You have no items in your shopping cart.
The UviLux is an innovative, sensitive, low cost, digital in-situ UV fluorimeter that offers the user significantly increased functionality compared to existing UV fluorimeters. It can monitor in real time the concentration of refined hydrocarbons (365nm) or crude hydrocarbons (450nm) in a wide range of coastal, oceanographic and fresh water applications. It has been designed into an acetal housing suitable for deploying independently as well as from submersible vehicles, moored or profiling systems. This robust, compact, lightweight fluorimeter has very low power consumption, is easy to use and gives accurate and repeatable measurements.
The fluorometer utilizes a novel UV LED light source and a miniature 8-stage photomultiplier coupled to a low noise preamp, when combined with phase sensitive detection this provides highly sensitive measurement capabilities. The sophisticated electronic signal processing, combined with cowl design principles employed in our UV AQUAtracka, enables the UviLux to operate successfully in high levels of ambient light. Because of this, there is no requirement to pump sample through a dark measurement chamber, thus obviating the need for water flow corrections or the cost, inconvenience and power drain of a pump. For deck & laboratory applications, a flow through manifold is available.
Ease of use and installation is further enhanced by the provision of galvanic isolation between the power and signal 0V rails within UviLux, thereby avoiding possible 'ground loops'. The highly efficient, proprietary, d.c. converter can also handle an unusually wide range of supply voltages ranging between 9V and 36V without significant change in the power consumption of 1.3W.
Two signals can be obtained from the standard UviLux: the first is a digital RS232 output in engineering units, the second a 0V to 5V volt calibrated analogue output derived from the digital signal. A single RS422 output is also available as an option. This flexibility makes the UviLux suitable for integrating into many different systems and platforms.
A Windows based Graphical User Interface (GUI) is provided that allows the user to both plot and record time stamped data when operating the UviLux directly from a PC and gives control over a many instrument parameters, including: sampling rate, detector sensitivity, and calibration factors. Although small and low cost, the UviLux fluorimeter maintains the performance standards and build quality previously maintained by the CTG’s range of in-situ fluorimeters.
Performance | | |
| |
Parameter | Refined Hydrocarbon | Crude Hydrocarbon |
| |
Excitation Wavelength: | 255nm | 255nm |
| |
Emission Wavelength: | 365nm | 450nm |
| |
Calibrated range: | 0 to 10μg/L carbazole | 0 to 10μg/L perylene |
| |
Limit of detection: | <0.02μg/L | <0.02μg/L |
| |
Accuracy: | <0.04 | <0.04 |
| |
Mechanical | ||||
Size: | 70mm dia x 149mm | |||
Weight in air: | 0.8 kg | |||
Weight in water: | 0.15 kg | |||
Pressure Housing: | Acetal C | |||
Depth rating: | 600 metres | |||
Connector: | Subcon MCBH6M | |||
Environmental | ||||
Operating Temperature: | -2ºC to +40ºC | |||
Storage Temperature: | -40ºC to +70ºC | |||
Electrical | ||||
Input voltage: | 9 to 36Vdc | |||
Data output: | Digital RS232 and analogue 0 to 5Vdc (RS422 as an option) | |||
Power requirements: | <1Watt @ 12 volt | |||
Delivery Package | ||||
The standard UviLux fluorimeter is supplied with transit case, handbook and Windows based Graphical User Interface (GUI).Accessories: User to specify cables and other accessories. | ||||
Applications
- In-situ hydrocarbon detection / monitoring of crude and refined oil.
- Hydrocarbon detection around drilling rigs
- Oil spill monitoring
- Sub-sea pipeline leak detection
- Airport apron pollution runoff
- Coastal pollution monitoring
- Environmental impact assessment
- Biomass indicator studies (CDOM)
- Designed for use on profiling, towed, moored or ROV deployments
- Biogeochemical oceanography
Features:
- Small, low cost in-situ fluorimeter
- Range of wavelengths available
- Digital output in engineering units
- Additional analogue output as standard
- User selectable sampling rate, 0.1Hz - 3Hz
- Low power consumption
- High rejection of ambient daylight
- Low noise, high sensitivity
- Low turbidity breakthrough
- User adjustable dynamic range
- RS232 & analogue output (or RS422 option)
- Acetal C housing
The standard UviLux fluorimeter is supplied with transit case, handbook and Windows based Graphical User Interface (GUI). There is now a range of accessories available to compliment the UviLux Fluorimeter. You will need to decide on cable type, length and cable termination. You will also need to give consideration to how you wish to interface the fluorimeter to either a laptop, PDA or your own system.
To help you decide which accessories you may need for the UviLux fluorimeters, please Click here.
If you need advice regarding which accessories you may need, please telephone Ellen Keegan on +44(0)20 8481 9019. Email: ekeegan@chelsea.co.uk.